Design and performance of a cesium iodide detectorстатья
Информация о цитировании статьи получена из
Scopus,
Web of Science
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 18 июля 2013 г.
Аннотация:The design, construction, and performance of a 198-element CsI detector built for Brookhaven experiment E852 is described. Design considerations for the array included such factors as rate, magnetic field, sensitivity and acceptance. Signals were obtained with a photodiode/preamplifier combination using PIN photodiodes. Data were taken over the course of two runs during the summers of 1993 and 1994. A calibration procedure using halo muons is described. The gain, energy resolution, and position resolution of the detector are discussed. Finally, the ability of the detector to be used as a low energy photon veto is illustrated using the data.