Atomic Absorption Determination of Elements in Solid Samples Using an Electrothermal Atomizer ‘Crucible with Separated Zonesстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 28 января 2015 г.
Аннотация:A model of an electrothermal atomizer given the name ‘crucible with separated zones’ has been
proposed. It provides the filtration of sample vapors through porous walls of a graphite tube placed in the atomization zone and heated independently of the evaporation zone. The inner volume of the tube is an isothermal analytical zone. The atomizer has been applied to the analysis of solid sea and river suspensions,DETATA concentrates of waters, and bottom sediments weighing 30–50 mg and more. As a result, the performance characteristics of the direct atomic absorption determination of Cd, Ag, and Pb in geochemical samples of complicated composition have been improved.